Limited application in the acute monitoring of heparin therapy during cardiopulmonary surgery and in intensive care and dialysis units.
Available methods have variable and limited correlation with the APTT and may be influenced by other variables (eg, PCV and fibrinogen).
A specific instrument should only be selected for use if its performance has been validated against the methods used in the pathology laboratory. Correlation with the APTT and plasma heparin levels should be checked at regular intervals.
|